Supply of pressurized ion chamber survey meter, as per the detail of this specification.
Technical specifications for Supply of Pressurized Ion Chamber Survey Meter
Quantity: 3 Systems
Functional Requirement
The pressurized ionization chamber survey meter shall be used for conducting radiation survey around an electron accelerator based facility. The survey meter should measure both exposure rate and integrated exposure simultaneously and should have auto ranging provision for measurement in the entire range as specified below.
Technical Specifications
S.No.Parameter Tender specification
3.1Radiation detected
Gamma and X-rays> 25 keV
3.2DetectorPressurized ion chamber
Nominal pressure 125 PSI
3.3Operating ranges (auto ranging)0-500 uR/h
0-5 mR/h
0-50 mR/h
0-500 mR/h
0-5 R/h
3.4Accuracy? 10% of the reading
3.5Precision? 5 %of the reading
3.6Response time<10s
3.7Automatic featureAuto zeroing and Auto ranging
3.8Warm -up time< 5 min
3.9DisplayAnalog : Bar graph and Digital : LCD
3.10Eriergy dependency? 20 % at 40 keV X-rays
3.11Mode of operationRate mode, Integrate mode and Freeze mode
3.12Power requirementBattery operated
3.13Temperature range20-50 "C?
3.14Humidity0-90 %
3.15Weight< 2kg
3.16Calibration traceabilityThe equipment should be supplied with.
NIST/NPL traceability calibration certificate